Technical Data Metrax XD 100
| Impulse shape | biphasic, current-controlled impulses, external |
| Mode Energy in the automatic mode | 140 - 360J |
| Manual energy stages | 5 J, 10 J, 20 J, 30 J, 50 J, 100 J, 200 J, 300 J, 360 J |
| Charge time | < 6 sec. for 200J |
| Analysis time | 7-12 s |
| Asystole threshold | < 200 V |
| Artefact detection | |
| Pacemaker detection | |
| Cardioversion | Manual activation, QRS Marker |
| ECG | |
| Derivation | Eindhoven I, II, III Goldberger avL, avR, aVF |
| Patient resistance | 23 - 200 Ohm |
| Heart rate | 30-300 per min |
| Alarms | 30-300 per min |
| Monitor | |
| Model | TFT, LCD, blue mode |
| Dimensions | 115 x 86 mm (diagonal 5,7"/144 mm) |
| Resolution | 320 x 240 pixels |
| Defibrillation electrodes | |
| Adhesive electrodes (hands free) | |
| Shelf life | 36 months |
| Cable length | 3,5 m SavePads Connect |
| Active electrode surface | 164 cm2 overall |
| Paddles | With built-in paediatric electrodes |
| Printer | |
| Model | Thermo-Direktverfahren / Thermal direct process |
| Resolution | 200 dpi |
| Paper width | 58 mm |
| Print width | 50 mm |
| Advance | 25 mm/s / 50 mm/s |
| Operating modes | AUTO, ONLINE, MEMO print |
| Event marking | Printout via MEMO print |
| Data management | |
| Removable memory | CF card 512 MB |
| Data/voice recording | > 90 h data and voice recording |
| Voice recording can be activated / deactivated | |
| Read out software | |
| Data management | |
| Removable memory | CF card 512 MB |
| Power supply | |
| AkuPak LITE | |
| Model | LiFePO4 |
| Capacity (nominal) | 2,3 Ah |
| Discharges/shocks | 150@200J |
| Monitoring | 5 h |
| Charge time | approx. 3 h |
| Built-in power supply | |
| Voltage | 90 V ... 264 V / 50 Hz / 60 Hz |
| Battery charging | Possible in the device |
| Resuscitation guidelines | AHA and ERC at factory |
| Dimensions | 33 cm x 29 cm x 16 cm |
| Weight incl. power supply | approx. 5.9 kg |